The first studies on ethidium bromide (EtBr) for nucleic acid quantification date back to the mid-1960s (1) and have been followed by occasional reports over the years (2,3). Conventional knowledge on spectrofluorimetric DNA quantification using EtBr indicates that the best results are obtained with 5 µg/mL EtBr, using excitation at 546 or 302 nm. This allows reaching a detection limit of 100 ng/mL of double-stranded DNA (2). Our studies indicate that these parameters are largely suboptimal, and that a significantly higher sensitivity can be obtained by lowering the concentration of EtBr and using a more efficient excitation in the ultraviolet (UV) region. We have also developed a novel ratiometric approach that allows to further improve the sensitivity of the assay.
To reach the best signal-to-noise ratios, EtBr (Sigma-Aldrich, St. Louis, MO, USA) was tested at different concentrations (0.1, 0.5, and 2.5 µg/mL), as previously done for Hoechst 33258 (2). EtBr was diluted in Tris-EDTA (TE) buffer (1 mM Tris-HCl, 0.1 mM EDTA, pH 7.4) in a final volume of 200 µL/well. Spectrofluorimetry microplates (OptiPlate™-96 white opaque 96-well microplate; PerkinElmer, Waltham, MA, USA) were utilized for all the assays presented here. However, conveniently, almost equivalent results were obtained using standard tissue culture flat-bottom 96-well plates (BD Falcon™ 96-well microplates; BD Biosciences, Franklin Lakes, NJ, USA). Genomic DNA from peripheral blood leukocytes (0.24 ng/well to 2 µg/well) prepared as previously described (4) was added to the wells, and fluorescence was measured immediately thereafter. Fluorescent readings were performed on a SpectraMax® Gemini XS spectrofluorimeter (Molecular Devices, Sunnyvale, CA, USA).
In Figure 1 we show the UV region excitation spectra of free and DNA-bound EtBr and demonstrate the range of linear measurements of DNA concentrations. Excitation peaks were at 286 and 270 nm for free and DNA-bound ethidium, respectively (Figure 1A). Highest emission was at 605 nm. We found the highest fluorescence increase of DNA-bound versus free EtBr (about 40-fold) and the best signal-to-noise ratio at 250 nm (Figure 1B and data not shown). A corresponding ratio of bound versus free signal was observed at 364-nm excitation, but at the price of a lower signal-to-noise ratio. On the other hand, we observed a significantly worse performance when exciting at 546 nm (2). Excitation at 250 nm/emission at 605 nm were chosen for all subsequent studies. This wavelength choice minimized chances of undesired bleed-through of excitation light through the diffraction grating monochromator, which can be particularly relevant when excitation is close to half of the emission wavelength chosen. Additionally, a 590-nm long-pass filter was utilized before light collection by the photomultiplier.
A concentration of 0.5 µg/mL EtBr allowed to reach a detection limit of 2 ng DNA/well (10 ng/mL), with a linear range of measurements from 4-250 ng/well (20–1250 ng/mL) (Figure 1C). The standard error of the mean (sem) of five independent experiments (Figure 1C) was shown to range between 0.4% (125–250 ng/well) and 2.2% (4–8 ng/well) of the respective means. Higher concentrations of EtBr extended the linear range of measurements to higher amounts of DNA (from 15 ng DNA/well to 1250 ng/well for 2.5 µg/mL EtBr). However, this was associated with a considerable loss in sensitivity (data not shown). On the other hand, sensitivity was slightly improved using 0.1 µg/mL EtBr (1 ng DNA/well; 5 ng/mL), but, the amplitude of the linear range was significantly reduced (2 to 50 ng/well; Figure 1D). Hence, for most purposes a good compromise between sensitivity and amplitude of the linear range can be obtained with 0.5 µg/mL EtBr. Fluorescence measurements on different dilution series of a specific DNA sample were found to be remarkably reproducible. Equally reproducible were fluorescence measurements on different or independently quantified sources of DNA (Figure 1 and data not shown).
As DNA binding induces both a fluorescence increase and a spectral shift of ethidium fluorescence (the excitation peak shifts from 286 to 270 nm), we also investigated if a ratio-metric approach could increase the sensitivity of EtBr-based measurements. Ratiometric methods are commonly used in spectrofluorimetry, as they provide a stringent internal standard, leading to more accurate measurements of both quantitative and qualitative fluorescent changes. Plotting the ratio between the fluorescent emissions of EtBr excitated at 250 nm versus 286 nm, we extended the linear range of measurements to lower DNA amounts and obtained a significant gain in sensitivity (2- to 4-fold) at any EtBr concentration tested. The lowest detection limit of ratiometric measurements was 0.25 ng DNA/well with 0.1 µg/mL EtBr (Figure 1E), with a linear range of 0.5-8 ng/well (2.5–40 ng/mL).
Parallel measurements were performed using quartz cuvettes (SUPRASIL® quartz cuvettes; Hellma GmbH & Co, Müllheim, Germany) and a PerkinElmer LS 45 Luminescence Spectrometer. As expected, somewhat better measurements were obtained, due to the better light path (i.e., absence of meniscus and reduced light scattering). However, the improvement in sensitivity was found marginal (<2-fold). The R2 (correlation coefficient) of serially diluted DNA in plate readings was found to be 0.996 (Figure 1C, inset); that of the corresponding cuvettes readings was a comparable 0.993. These findings and the more cumbersome handling of DNA-EtBr dilutions discouraged the use of cuvettes when several dozen samples had to be measured. It should be noted that disposable plastic cuvettes cannot be used for excitation in the UV region around 250 nm. In this case, excitation at 364 nm is advisable.
Fluorescence measurements on different dilution series of a specific DNA sample and across different DNA samples, all previously quantified by spectrophotometry, were found to be highly reproducible. Fluorescence measurements closely matched the expected amount of DNA contained in a well down to 0.5 ng DNA/well (0.1 µg/mL EtBr; Figure 1E). The shape of the excitation spectra of DNA/EtBr samples were as expected from mixtures of the two corresponding spectra at different ratios (Figure 1B).
In summary, our data render better justice to EtBr, a fluorescent dye long forgotten in spectrofluorimetric DNA quantification. We found that the choice of appropriate EtBr concentrations and better excitation/emission wavelengths, as well as the use of ratiometric measurements make the EtBr-based spectrofluorimetric DNA quantification significantly more performing than currently thought (3,5,6).
The financial support of the Fondazione of the Cassa di Risparmio della Provincia di Chieti, of the Association for the Application of Biotechnology in Oncology (ABO, grant no. VE01D0019), and of the Italian Ministry for the University and Research (FIRB Postgenomica, grant no. RBNE0157EH) is gratefully acknowledged.
The authors declare no competing interests.